Electrochemical detection of microcystins in water

Doctoral researchEnvironmental biosensingEIS + amperometry
gold working electrode on Si/SiO₂label-free immunosensing of microcystin-LRMC-LR[Fe(CN)₆]³⁻/⁴⁻EIS · NyquistRct ↑ with toxin
Monoclonal antibodies on a gold electrode capture microcystin; binding hinders a redox probe and raises the charge-transfer resistance measured by impedance spectroscopy. Schematic.
01

Research problem

Microcystins are liver toxins produced during cyanobacterial blooms, and microcystin-LR is the most common variant in surface water. Regulators set very low limits, yet standard methods such as ELISA and LC-MS need a laboratory, trained staff and days of turnaround. Water managers need a fast, portable measurement at the source.

02

Project objective

Develop miniaturised electrochemical immunosensors that quantify microcystin-LR below guideline levels, and evaluate them in treated wastewater against ELISA, not only in buffer.

03

My contribution

  • Fabricated the single and multiplexed gold-electrode chips in the cleanroom
  • Developed the surface chemistry: monolayers, antibody and antigen immobilisation, blocking
  • Designed and optimised both assays and the electrochemical protocols, including equivalent-circuit models
  • Planned the validation: specificity controls, ELISA comparison and real samples
  • Wrote three first-author papers and co-authored the patent application
04

Methods and experimental approach

  • Label-free assay: electrochemical impedance spectroscopy with monoclonal anti-microcystin antibodies
  • Enzyme-assisted assay: competitive, enzyme-labelled amperometric immunoassay on the same chip design with toxin immobilised by EDC/NHS chemistry
  • Specificity: toxin-producing Microcystis aeruginosa, non-producing Spirulina and non-specific antibody controls
  • Real samples: treated-wastewater samples from secondary-effluent reservoirs, benchmarked against ELISA
05

Key findings

  • Label-free EIS: detection limit 0.003 µg/L (3 ng/L), dynamic range 0.003–30 µg/L (five orders of magnitude), measurement in under 10 minutes
  • Enzyme-assisted amperometry: detection limit 0.066 µg/L, assay time about 20 minutes
  • Both methods agreed well with ELISA in treated wastewater below the WHO guideline of 1 µg/L
  • The EIS assay was more stable and reproducible; the amperometric assay gave an independent check at low concentrations

Limitations

  • The amperometric assay depends on enzyme activity and saturates at higher concentrations
  • Evaluation covered treated-wastewater samples; other water types and long-term field use were not tested
06

Images and data

Abraham Gabriel with a biosensor chip
Holding a microfabricated biosensor chip next to cyanobacteria cultures, with an impedance (Nyquist) measurement on screen.
Label-free impedimetric immunosensor for microcystin-LR Antibodies on a microfabricated gold chip capture microcystin-LR from surface water that also contains cyanobacteria. Binding changes the electrode impedance, read out electronically. surface-water sample Vout R impedance readout MC-LR anti-MC-LR antibody cyanobacteria Label-free EIS immunosensor · LOD 3 ng/L · five orders of magnitude · surface water
The label-free impedimetric immunosensor for surface water. Adapted from the graphical abstract of Ogungbile et al., Sensors and Actuators B: Chemical 348, 130687 (2021). Redrawn illustration.
Chip-based impedimetric detection compared with ELISA in treated wastewater Samples from secondary-effluent reservoirs were measured in parallel by a portable impedimetric biochip, read as a Nyquist plot, and by indirect competitive ELISA in a 96-well plate reader. The two methods gave statistically equivalent results. secondary-effluent reservoirs 24 water samples Impedimetric biosensor portable · label-free · minutes Indirect competitive ELISA laboratory · 96-well plate · hours portable potentiostat electrochemical biochip Nyquist 96-well pipetting plate reader ELISA curve statistically equivalent results Real treated wastewater · chip-based impedimetric sensor benchmarked against ELISA
Validation in real treated wastewater against ELISA. Adapted from the graphical abstract of Ogungbile et al., ACS ES&T Water 5(2), 696–702 (2025). Redrawn illustration.
Multi-method electrochemical biosensor for microcystins Left: label-free impedimetric detection, where microcystin binding to antibodies raises charge-transfer resistance in the Nyquist plot. Right: competitive amperometric detection, where free toxin blocks enzyme-labelled antibodies so less current is produced. Both run on the same chip type and were benchmarked against ELISA in treated wastewater. Label-free impedimetric antibodies capture the toxin directly Competitive amperometric free toxin blocks enzyme-labelled antibodies gold working electrode gold working electrode MC-LR substrate → product e⁻ Z′ (Ω) −Z″ Rct grows with [MC-LR] time (s) I I(t) = nFAC√D / √(πt) more free toxin → less current Same chip · two independent readouts · LOD 0.003 µg/L (EIS) and 0.066 µg/L (amperometry) · <10 / 20 min · benchmarked against ELISA
The two detection modes of the later multi-method biosensor. Redrawn from the graphical abstract of Ogungbile & Vernick, J. Electrochem. Soc. 172, 097503 (2025), open access under CC BY 4.0.
07

Publications and related outputs

  • Rapid detection and quantification of microcystins in surface water by an impedimetric immunosensor. Sensors and Actuators B: Chemical, 2021. DOI ↗
  • Real-time chip-based impedimetric detection of cyanotoxins in treated wastewater. ACS ES&T Water, 2025. DOI ↗
  • A Comparative Study of a Multi-Method Electrochemical Biosensor for Rapid Detection of Microcystins in Treated Wastewater. Journal of The Electrochemical Society, 2025. DOI ↗
  • Patent application: integrated multi-method electrochemical biosensor. PCT/IL2021/051409 · US 2025/0264433 A1