Doctoral · published
Electrochemical detection of microcystins in water
01
Research problem
Microcystins are liver toxins produced during cyanobacterial blooms, and microcystin-LR is the most common variant in surface water. Regulators set very low limits, yet standard methods such as ELISA and LC-MS need a laboratory, trained staff and days of turnaround. Water managers need a fast, portable measurement at the source.
02
Project objective
Develop miniaturised electrochemical immunosensors that quantify microcystin-LR below guideline levels, and evaluate them in treated wastewater against ELISA, not only in buffer.
03
My contribution
- Fabricated the single and multiplexed gold-electrode chips in the cleanroom
- Developed the surface chemistry: monolayers, antibody and antigen immobilisation, blocking
- Designed and optimised both assays and the electrochemical protocols, including equivalent-circuit models
- Planned the validation: specificity controls, ELISA comparison and real samples
- Wrote three first-author papers and co-authored the patent application
04
Methods and experimental approach
- Label-free assay: electrochemical impedance spectroscopy with monoclonal anti-microcystin antibodies
- Enzyme-assisted assay: competitive, enzyme-labelled amperometric immunoassay on the same chip design with toxin immobilised by EDC/NHS chemistry
- Specificity: toxin-producing Microcystis aeruginosa, non-producing Spirulina and non-specific antibody controls
- Real samples: treated-wastewater samples from secondary-effluent reservoirs, benchmarked against ELISA
05
Key findings
- Label-free EIS: detection limit 0.003 µg/L (3 ng/L), dynamic range 0.003–30 µg/L (five orders of magnitude), measurement in under 10 minutes
- Enzyme-assisted amperometry: detection limit 0.066 µg/L, assay time about 20 minutes
- Both methods agreed well with ELISA in treated wastewater below the WHO guideline of 1 µg/L
- The EIS assay was more stable and reproducible; the amperometric assay gave an independent check at low concentrations
Limitations
- The amperometric assay depends on enzyme activity and saturates at higher concentrations
- Evaluation covered treated-wastewater samples; other water types and long-term field use were not tested
06
Images and data

07
Publications and related outputs
- Rapid detection and quantification of microcystins in surface water by an impedimetric immunosensor. Sensors and Actuators B: Chemical, 2021. DOI ↗
- Real-time chip-based impedimetric detection of cyanotoxins in treated wastewater. ACS ES&T Water, 2025. DOI ↗
- A Comparative Study of a Multi-Method Electrochemical Biosensor for Rapid Detection of Microcystins in Treated Wastewater. Journal of The Electrochemical Society, 2025. DOI ↗
- Patent application: integrated multi-method electrochemical biosensor. PCT/IL2021/051409 · US 2025/0264433 A1