Two electrochemical readouts for microcystin detection: why compare them?

Multi-method electrochemical biosensor for microcystins Left: label-free impedimetric detection, where microcystin binding to antibodies raises charge-transfer resistance in the Nyquist plot. Right: competitive amperometric detection, where free toxin blocks enzyme-labelled antibodies so less current is produced. Both run on the same chip type and were benchmarked against ELISA in treated wastewater. Label-free impedimetric antibodies capture the toxin directly Competitive amperometric free toxin blocks enzyme-labelled antibodies gold working electrode gold working electrode MC-LR substrate → product e⁻ Z′ (Ω) −Z″ Rct grows with [MC-LR] time (s) I I(t) = nFAC√D / √(πt) more free toxin → less current Same chip · two independent readouts · LOD 0.003 µg/L (EIS) and 0.066 µg/L (amperometry) · <10 / 20 min · benchmarked against ELISA
Redrawn from the graphical abstract of Ogungbile & Vernick, J. Electrochem. Soc. 172, 097503 (2025), open access under CC BY 4.0.

Developing an electrochemical biosensor involves more than achieving a sensitive response under controlled laboratory conditions. It is also important to understand how consistently the method performs, and whether its results remain reliable in relevant samples.

During my doctoral research I explored two approaches for detecting microcystin-LR, a liver toxin released during cyanobacterial blooms: label-free electrochemical impedance spectroscopy (EIS) and enzyme-assisted amperometry. Both were developed on similar microfabricated biochips, but they differ in detection principle, sensitivity and assay time.

Label-free impedance

Antibodies on a gold electrode capture the toxin; the layer becomes harder for a redox probe to cross, and the charge-transfer resistance rises with concentration. No labels are needed, the detection limit was 3 ng/L, and a measurement takes under ten minutes. Treated wastewater, however, carries salts, organic matter and microorganisms that can also affect an impedance signal.

Enzyme-assisted amperometry

The second assay is competitive: toxin immobilised on the electrode competes with toxin in the sample for an enzyme-labelled antibody, and the enzyme's product is read as a current. It uses the same chip design but a different detection principle.

The two methods did not behave identically, and that was the useful part. The impedance sensor was more stable and reproducible, because it reads the binding directly. The amperometric assay depends on enzyme activity, which varies, and it saturates at high concentrations. At low concentrations, where regulation applies, it gave an independent check, with a detection limit of 0.066 µg/L in about 20 minutes. Across treated wastewater samples, results agreed well with ELISA below WHO guideline levels.

What the comparison showed

Comparing the two methods clarified what each can offer for water-quality monitoring: label-free EIS for sensitivity and speed, and amperometry as a second, mechanistically different measurement at low concentrations. It also showed why I plan controls, reference methods and testing in relevant samples early in a project.

The full comparison is in the Journal of The Electrochemical Society (2025). Read the project page →